Dynamic<RangePcl>

RepCap Settings

# Range: Nr1 .. Nr5
rc = driver.configure.multiEval.limit.qam.powerVsTime.upper.fallingEdge.dynamic.repcap_rangePcl_get()
driver.configure.multiEval.limit.qam.powerVsTime.upper.fallingEdge.dynamic.repcap_rangePcl_set(repcap.RangePcl.Nr1)

SCPI Commands

CONFigure:GSM:MEASurement<Instance>:MEValuation:LIMit:QAM<QamOrder>:PVTime:UPPer:FEDGe<FallingEdge>:DYNamic<RangePcl>
class Dynamic[source]

Dynamic commands group definition. 1 total commands, 0 Sub-groups, 1 group commands Repeated Capability: RangePcl, default value after init: RangePcl.Nr1

class DynamicStruct[source]

Structure for setting input parameters. Fields:

  • Enable: bool: OFF | ON Disable or enable dynamic correction

  • Pcl_Start: float: numeric First PCL in PCL range Range: 0 to 31

  • Pcl_End: float: numeric Last PCL in PCL range (can be equal to PCLStart) Range: 0 to 31

  • Correction: float: numeric Correction value for power template Range: -100 dB to 100 dB, Unit: dB

get(qamOrder=<QamOrder.Default: -1>, fallingEdge=<FallingEdge.Default: -1>, rangePcl=<RangePcl.Default: -1>)DynamicStruct[source]
# SCPI: CONFigure:GSM:MEASurement<Instance>:MEValuation:LIMit:QAM<ModOrder>:PVTime:UPPer:FEDGe<nr>:DYNamic<Range>
value: DynamicStruct = driver.configure.multiEval.limit.qam.powerVsTime.upper.fallingEdge.dynamic.get(qamOrder = repcap.QamOrder.Default, fallingEdge = repcap.FallingEdge.Default, rangePcl = repcap.RangePcl.Default)

These commands define and activate dynamic (PCL-dependent) corrections to the upper limit lines for the measured power vs. time. The corrections apply to the modulation schemes GMSK, 8PSK (EPSK) or 16-QAM (QAM16) and to the three limit line sections: rising edge (REDGe) , useful part (UPARt) and falling edge (FEDGe) . Each limit line section consists of several areas (<no>) . Each dynamic correction is defined for up to five different PCL ranges (<Range>) ).

param qamOrder

optional repeated capability selector. Default value: Nr16 (settable in the interface ‘Qam’)

param fallingEdge

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘FallingEdge’)

param rangePcl

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Dynamic’)

return

structure: for return value, see the help for DynamicStruct structure arguments.

set(structure: RsCmwGsmMeas.Implementations.Configure_.MultiEval_.Limit_.Qam_.PowerVsTime_.Upper_.FallingEdge_.Dynamic.Dynamic.DynamicStruct, qamOrder=<QamOrder.Default: -1>, fallingEdge=<FallingEdge.Default: -1>, rangePcl=<RangePcl.Default: -1>)None[source]
# SCPI: CONFigure:GSM:MEASurement<Instance>:MEValuation:LIMit:QAM<ModOrder>:PVTime:UPPer:FEDGe<nr>:DYNamic<Range>
driver.configure.multiEval.limit.qam.powerVsTime.upper.fallingEdge.dynamic.set(value = [PROPERTY_STRUCT_NAME](), qamOrder = repcap.QamOrder.Default, fallingEdge = repcap.FallingEdge.Default, rangePcl = repcap.RangePcl.Default)

These commands define and activate dynamic (PCL-dependent) corrections to the upper limit lines for the measured power vs. time. The corrections apply to the modulation schemes GMSK, 8PSK (EPSK) or 16-QAM (QAM16) and to the three limit line sections: rising edge (REDGe) , useful part (UPARt) and falling edge (FEDGe) . Each limit line section consists of several areas (<no>) . Each dynamic correction is defined for up to five different PCL ranges (<Range>) ).

param structure

for set value, see the help for DynamicStruct structure arguments.

param qamOrder

optional repeated capability selector. Default value: Nr16 (settable in the interface ‘Qam’)

param fallingEdge

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘FallingEdge’)

param rangePcl

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Dynamic’)

Cloning the Group

# Create a clone of the original group, that exists independently
group2 = driver.configure.multiEval.limit.qam.powerVsTime.upper.fallingEdge.dynamic.clone()