Average

SCPI Commands

FETCh:GSM:MEASurement<Instance>:MEValuation:LIST:SEGMent<Segment>:PVTime:AVERage
CALCulate:GSM:MEASurement<Instance>:MEValuation:LIST:SEGMent<Segment>:PVTime:AVERage
class Average[source]

Average commands group definition. 3 total commands, 1 Sub-groups, 2 group commands

class CalculateStruct[source]

Response structure. Fields:

  • Reliability: int: decimal ‘Reliability Indicator’ In list mode, a zero reliability indicator indicates that the results in all measured segments are valid. A non-zero value indicates that an error occurred in at least one of the measured segments.

  • Seg_Reliability: int: decimal Reliability indicator for the segment. The meaning of the returned values is the same as for the common reliability indicator, see previous parameter.

  • Statist_Expired: int: decimal Number of measured steps Range: 0 to Statistical Length (integer value)

  • Slot_Info: enums.SlotInfo: No parameter help available

  • Slot_Statistic: bool: ON | OFF ON: Averaging over different burst type OFF: Uniform burst type in the averaging range

  • Out_Of_Tolerance: int: decimal Percentage of measured bursts with failed limit check Range: 0 % to 100 %, Unit: %

  • Average_Burst_Pow: float: No parameter help available

class FetchStruct[source]

Response structure. Fields:

  • Reliability: int: decimal ‘Reliability Indicator’ In list mode, a zero reliability indicator indicates that the results in all measured segments are valid. A non-zero value indicates that an error occurred in at least one of the measured segments.

  • Seg_Reliability: int: decimal Reliability indicator for the segment. The meaning of the returned values is the same as for the common reliability indicator, see previous parameter.

  • Statist_Expired: int: decimal Number of measured steps Range: 0 to Statistical Length (integer value)

  • Slot_Info: enums.SlotInfo: No parameter help available

  • Slot_Statistic: bool: ON | OFF ON: Averaging over different burst type OFF: Uniform burst type in the averaging range

  • Out_Of_Tolerance: int: decimal Percentage of measured bursts with failed limit check Range: 0 % to 100 %, Unit: %

  • Average_Burst_Pow: float: No parameter help available

calculate(segment=<Segment.Default: -1>)CalculateStruct[source]
# SCPI: CALCulate:GSM:MEASurement<Instance>:MEValuation:LIST:SEGMent<nr>:PVTime:AVERage
value: CalculateStruct = driver.multiEval.listPy.segment.powerVsTime.average.calculate(segment = repcap.Segment.Default)

Returns power vs. time results for segment <no> in list mode. The values described below are returned by FETCh commands. The first six values (‘Reliability’ to ‘Out of Tolerance’ result) are also returned by CALCulate commands. The remaining values returned by CALCulate commands are limit check results, one value for each result listed below.

param segment

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Segment’)

return

structure: for return value, see the help for CalculateStruct structure arguments.

fetch(segment=<Segment.Default: -1>)FetchStruct[source]
# SCPI: FETCh:GSM:MEASurement<Instance>:MEValuation:LIST:SEGMent<nr>:PVTime:AVERage
value: FetchStruct = driver.multiEval.listPy.segment.powerVsTime.average.fetch(segment = repcap.Segment.Default)

Returns power vs. time results for segment <no> in list mode. The values described below are returned by FETCh commands. The first six values (‘Reliability’ to ‘Out of Tolerance’ result) are also returned by CALCulate commands. The remaining values returned by CALCulate commands are limit check results, one value for each result listed below.

param segment

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Segment’)

return

structure: for return value, see the help for FetchStruct structure arguments.

Cloning the Group

# Create a clone of the original group, that exists independently
group2 = driver.multiEval.listPy.segment.powerVsTime.average.clone()

Subgroups